Description
EziLab supplies the UV-Visible Spectrophotometer EZL-US115 for pharmaceutical and research laboratories that need standard absorbance measurement across the full 190-1100nm range at 2nm bandwidth, covering both UV and visible-light testing on one instrument. This UV spectrophotometer supports the baseline wavelength and bandwidth combination most routine pharmaceutical and research absorbance work calls for, from raw material identification to general concentration determination, without requiring the finer resolution or specialized beam configurations found on other spectrophotometer models in this catalog. That 190-1100nm coverage spans the practical range most absorbance-based applications operate within, giving pharmaceutical QC labs and research programs a versatile instrument for everyday spectrophotometric testing across compound classes that read within either the UV or visible spectrum. The 2nm bandwidth balances resolution and scan speed for routine testing, adequate for most standard absorbance measurements without needing the tighter bandwidths reserved for closely spaced peak resolution work. Pharmaceutical and research labs running general-purpose absorbance testing across a variety of sample types get broad practical coverage from this standard configuration. See the UV Spectrophotometer and UV-Visible Spectrophotometer ranges for general-purpose absorbance testing needs across USA pharmaceutical and research laboratory settings.
Specifications
| Wavelength Range | 190 nm – 1100 nm |
|---|---|
| Spectral Bandwidth | 2 nm |
| Optical System | Single beam (1200 lines/mm grating) |
| Wavelength Accuracy | ±0.5 nm |
| Wavelength Repeatability | ±0.2 nm |
| Photometric Display Range | -0.3 – 3A, 0 – 200%T, 0 – 9999C |
| Photometric Accuracy | ±0.3%T |
| Photometric Repeatability | 0.2%T |
| Stability | ±0.002A/h @ 500 nm |
| Baseline Flatness | ±0.002A |
| Stray Light | ≤0.05%T @ 220 nm, 360 nm |
| Noise | ±0.001A |
| Cell Holder | 4 single cell holder |
| Display | LCD (320×240 Dots) |
| Lamps | Deuterium and Tungsten Halogen lamps |
| Detector | Silicon photodiode |
| Data Output Port | USB |
| Printer Port | Parallel port |
| Packing Dimension (W×D×H) | 70×61×44 cm |
| Gross Weight | 25 Kg |
Features