EziLab offers the Muffle Furnace EZL-MF101, a 1 L benchtop laboratory muffle furnace that heats a 100 mm cubic ceramic fiber chamber to 1200 °C, or 1100 °C continuously, with resistance wire elements. A digital PID controller holds the setpoint within ±1 °C from a K type thermocouple, ramps at up to 30 °C/min, and shows current and set temperatures on separate readouts. The low-mass fiber lining heats quickly on 1 kW, the double-layer steel casing with fan cooling keeps the outer surface below 50 °C, and a top vent carries off fumes during sintering. An over-temperature limiter and sensor alarm add protection. At 12 kg and 260 mm wide, the unit fits any lab bench. Materials science laboratories, dental technicians, ceramic and glass makers, metallurgy departments, analytical chemistry labs and university teaching facilities use this muffle furnace for high-temperature work. Typical uses include ashing of food, coal and polymer samples, loss-on-ignition tests, ceramic sintering, dental crown firing, heat treatment of small metal parts and annealing of glass. Choose it over larger chamber furnaces when quick cycles on gram-scale samples matter more than throughput. New fiber chambers and elements often give off odor and fine dust during the first firing; run that cycle empty under extraction, with Furnace and Muffle Furnace listings open for sizing.



Quantity:
$1790
Capacity 1 L
Maximum Working Temperature 1200°C
Continuous Working Temperature RT to 1100°C
Heating Rate 0 to 20°CMaximum Heating Rate: 30°C/min
Temperature Accuracy ±1°C
Thermocouple K type
Heating Element Resistance Wire
Temperature Control PID automatic control
Furnace Material Double layer steel casing
Rated Power 1 Kw
Working Voltage AC 110/208V-240V, 50/60Hz
Chamber Dimension (W×D×H) 100×100×100 mm
Overall Dimension (W×D×H) 260×400×280 mm
Net Weight 12 kg
Feature AreaDescription
ChamberA ceramic fiber chamber heats quickly and saves power.
CasingDouble layer steel casing with fan cooling keeps the outer surface cool.
ControllerA digital PID controller shows current and set temperature separately on the panel.
SafetyAn over-temperature limit function and a sensor alarm guard the run.
VentingA top vent hole carries waste gas away during material sintering.
  • Supports ashing and loss-on-ignition work.
  • Handles sintering of materials.
  • Suited to dental firing in dental labs.
  • Applied to heat treatment in metallurgy work.
  • Employed for glass annealing.
  • Fits ceramics and analytical chemistry settings.