Gas Chromatograph EZL-GC62 Catalog

Description

Gas chromatograph EZL-GC62 features a 22L oven capacity and a temperature range from RT up 5°C to 400°C. This triggers GC112A and workstation externally, compatible with devices like automatic samplers or thermal analyzers. It has 8 external event interfaces, expandable and compatible with various control valves and programmable timing. Our Gas Chromatograph includes a built-in self-check sound system and automatic malfunction identification.

Specifications

Oven Inside capacity 22L
Oven Temperature Range RT up 5°C to 400°C
Oven Temperature Accuracy ±0.1°C
Oven Temperature Programming 9 steps
Oven Total Program Time 9999.9 min
Oven Max Heating Rate 0.1 to 60°C /min
Oven Max Cooling Rate 10min (250°C to 50°C)
Sample Injector Temperature Range RT up 7°C to 420°C
Sample Injector Max Quantity 3PCS
Sample Injector Type Packed Column or Split
Sample Injector Split Ratio Display
Sample Injector Pre-Column Pressure Range 0 to 400 Kpa
Sample Injector Pre-Column Pressure Accuracy 0.1 Kpa
Sample Injector Flow Setting Range H2 0 to 200ml/min, N2 0 to 150ml/min
Detector FID or TCD
Detector Temperature Range RT up 7°C to 420°C
Detector Max installation Qty 2
Detector Ignition model Automatic
Hydrogen Flame Ionization Detector (FID) Logarithmic Amplifier, High Voltage Switch Control, Baseline Signal Display
FID Detector Detection limit ≤ 3x10⁻¹² g/s (Sample: n-hexadecane)
FID Detector Baseline noise ≤ 5x10⁻¹⁴A
FID Detector Baseline draft ≤ 6x10⁻¹³ A
FID Detector Dynamic range 10⁷
FID Detector RSD ≤ 3%
Thermal Conductivity Detector (TCD) Sensitivity: 8000mV.ml/mg (Sample: n-hexadecane)
TCD Detector Baseline noise ≤ 0.05mV
TCD Detector Baseline draft 10⁵
Memory Function Storage 20 Sample Test Modes
Power Suply 220V±22V, 50Hz±0.5Hz
Power Consumption 3000W
Overall Dimension 1020 × 660 × 740 mm
GrossWeight 82kg
Net Weight 70kg

Features

  • ✓ Convenient one-key startup function
  • ✓ RS232 and LAN ports for connectivity
  • ✓ Reliable multi-core, 32-bit hardware system
  • ✓ GC112A offers bidirectional touch screen control
  • ✓ Constant carrier gas control flow
  • ✓ Independent temperature control
  • ✓ 7-inch LED color display

Standard Accessories

  • ✓ FID Detector

Optional Accessories

  • ✓ Chromatography Workstation Software

Applications

  • ✓ Gas chromatography finds extensive use in laboratory research, research and development, and pharmaceutical quality control, enhancing efficiency, accuracy, and reliability in analytical and research efforts.