Gas Chromatography Mass Spectrometry EZL-GCMS100 Catalog

Description

EziLab offers the Gas Chromatography Mass Spectrometry EZL-GCMS100, a single quadrupole GC-MS system that separates volatile and semi-volatile compounds and identifies them by mass spectrum in one run. Choose it when GC with FID or ECD can quantify targets but cannot confirm what an unknown peak is. The quadrupole covers 1.5 to 1000 amu at unit mass resolving power, scans at up to 10,000 amu/s for fast peaks, and holds mass stability to ±0.10 amu over 48 hours. An EI source adjustable from 5 to 250 eV and 100 to 350°C, a high energy dynode electron multiplier, and a 67 L/s turbomolecular pump deliver full scan sensitivity of S/N ≥ 30:1. The GC oven runs from 4 to 450°C at up to 120°C/min across 7 ramps and 8 hold steps, with electronic pressure control in CV and CC modes from 0 to 100 psi, and split/splitless injection up to 1000:1. Pharmaceutical labs test residual solvents and trace impurities, environmental agencies screen VOCs and contaminants, forensic and food safety teams confirm drug residues, petrochemical units characterize complex mixtures, and semiconductor labs check wafer and plastic extract contamination. See the GC MS and Gas Chromatography Mass Spectrometry EZL-GCMS100 pages for full specifications.

Specifications

Gas Chromatography Flow Rate 1 to 1.5 ml/min
Maximum Flow Rate 10 ml/min
Inlet Temperature 450 °C
Pressure Range 0 to 100 psi
Heating Rate Up to 120 °C / min
Temperature Range 4 °C to 450 °C
Pressure Control Mode Electronic pressure control ( Supports CV and CC )
Split Mode Split/split less
Split Ratio 1000:1
Temperature programming 7 stages / 8 platforms
Autosampler Optional
Mass Spectrometry Mass Range 1.5 to 1000 amu
Ion – Source Temperature 100 °C to 350 °C
GC-MS Interface Temperature 450 °C
Scan Rate Up to 10000 amu/s
Stability ± 0.10 amu / 48 hrs
Filament Emission Current 0 to 350 µA
EI Source Ionization Energy 5 eV to 250 eV
Resolution Unit Resolution
Detector High energy dynode electron multiplier
Sensitivity Full scan ( S / N is ≥ 30 : 1 )
Vacuum Turbo molecular pump (67 L/s)

Applications

  • ✓ Gas Chromatography Mass Spectrometry EZL-GCMS100 is utilized for identifying and quantifying volatile organic compounds, testing residual solvents, and detecting trace impurities in liquids, gases, plastic extracts, and contaminants on semiconductor wafers, and other advanced materials.