Description
EziLab Portable Water Quality Analyzer EZL-PWA10 performs multi-parameter analysis including COD, ammonia nitrogen, total phosphorus, and total nitrogen across aquaculture monitoring, wastewater treatment evaluation, river and lake surveys, industrial discharge checks, and environmental compliance testing applications. LED light source maintains consistent wavelength stability during water quality analyzer on-site testing, while compact handheld format enables fast analysis without laboratory ency. Built-in data storage records multiple test results for later review, and short digestion cycles improve testing turnaround time. Battery-powered operation supports mobility across diverse sampling locations without infrastructure requirements. EziLab's Lab Equipment and water quality analyzers unite the complete water quality analyzer and portable multi-parameter testing range for field professionals.
Specifications
| COD Test Method | Potassium Dichromate |
|---|---|
| COD Measuring Range | 10 to 15000 mg/L |
| COD Test Limit | 10 mg/L |
| COD Digestion Temperature | 165 ˚C, 20 minutes |
| Ammonia Nitrogen Test Method | Salicylic acid method |
| Ammonia Nitrogen Measuring Range | 0.02 to 50 mg/L |
| Ammonia Nitrogen Test Limit | 0.02 mg/L |
| Ammonia Nitrogen Digestion Temperature | NA |
| Total Phosphorus Test Method | Ammonium Molybdate |
| Total Phosphorus Measuring Range | 0.02 to 20 mg/L |
| Total Phosphorus Test Limit | 0.02 mg/L |
| Total Phosphorus Digestion Temperature | 150 ˚C, 15 minutes |
| Total Nitrogen Test Method | Thymol Crystals |
| Total Nitrogen Measuring Range | 0.5 to 500 mg / L |
| Total Nitrogen Test Limit | 0.5 mg/L |
| Total Nitrogen Digestion Temperature | 125 ˚C, 30 minutes |
| Light Source | LED |
| Wavelength | 420nm / 520nm/ 620nm |
| Accuracy | ≤ ± 5% |
| Repeatability | ≤ ± 3% |
| Storing Data | 300 seats |
| Zero Drift | ≤ 0.3 % F.S/ 30 minutes |
| Dimension | 90 × 70 × 125mm |
| Weight | 300g |
Features
Applications