EziLab offers the Sample Concentrator EZL-SC123, a nitrogen blowdown evaporator that pairs a heated dry block with gas needles to concentrate many samples at once. The block heats from RT+5°C to 160°C, reaching 160°C from 25°C in 15 minutes or less, with ±0.5°C uniformity and 0.1°C display resolution. Temperature holds within ±0.5°C at 40°C and ±1°C at 120°C. A timer runs from 1 minute to 99 h 59 min or continuously. The 200 mm needles deliver up to 15 L/min of gas at a maximum of 0.1 MPa, blowing a gentle stream across each sample surface to speed solvent removal. The compact 260 × 220 × 450 mm, 6 kg unit draws 300 W at 110 V or 220 V. As a nitrogen evaporator, it dries and concentrates extracts before LC, LC-MS, and GC-MS analysis, so chemists can reconstitute residues in smaller volumes and lift detection limits. Toxicology and forensic labs prepare drug screening samples, and clinical labs concentrate steroid hormone extracts. Commodity inspection and food labs prepare residue and contaminant extracts, environmental labs process water and soil extracts, and pharmacy and biological product labs concentrate samples during development. See the Sample Concentrators and Sample Concentrator pages for full specifications.



Temperature Range RT+ 5°C to 160°C
Temperature Accuracy (@40 ℃) ± 0.5℃
Temperature Accuracy (@120 ℃) ±1℃
Temperature Uniformity ± 0.5℃
Display Accuracy 0.1℃
Time Range 1 to 99 h 59min/∞
Heating Time ≤ 15min(from 25℃ to 160℃)
Needle Length 200mm
Maximum Gas Usage 15 L/min
Maximum Gas Pressure 0.1 Mpa
Block Quantity 1
Maximum Power 300W
Voltage AC220V or AC110V/50 to 60Hz
Dimension ( W x L x H ) 260 x 220 x 450mm
Weight 6.0kg
Feature AreaDescription
Blowdown MethodNitrogen blowdown over a dry block heater speeds solvent removal from samples.
Batch ProcessingMany samples can be blown down at the same time.
Sample BlockSingle heated block holds the sample tubes in place.
Gas DeliveryAdjustable gas flow reaches the samples through needles.
Timer & ControlDigital temperature setting with a timer and a continuous run mode.
  • Supports analytical labs preparing samples for LC, LC-MS and GC-MS.
  • Handles drug screening sample preparation.
  • Suited to hormone analysis labs evaporating extracts at gentle temperatures.
  • Applied to food and environmental residue testing.
  • Employed by commodity, pharmacy and biological product labs running batches.