Description
EziLab offers the Bench Top pH Meter EZL-PM212, a high accuracy digital pH meter reading -2.00 to 19.99 pH in 0.01 steps and 1 mV on the millivolt scale, with drift held to ±0.05 pH plus one digit over three hours. Temperature compensation runs automatically from 0 °C to 80 °C or by hand, the firmware recognizes 4.00, 6.86 and 9.18 standard buffers, and calibration takes up to three points. A 10^12 Ω input stage keeps electrode loading negligible, Auto-Hold freezes a settled reading, a backlit LCD shows pH, mV and temperature together, and a reset key restores defaults. The 290 by 210 by 95 mm body weighs 1.5 kg and runs from 12 V DC. Choose it over a standard 0 to 14 pH unit when samples sit outside that range, such as concentrated acids and strong caustics. Pharmaceutical QC labs use it for release testing, water treatment plants check dosing baths, agricultural stations read soil extracts, and teaching labs run student titrations. Daily tasks cover buffer prep, plating bath control, and effluent pH logging. Strong caustics push ordinary glass electrodes into alkali error and wear them out sooner; ask which electrode type suits your highest-pH samples before ordering one. Lab Equipment and Laboratory pH Meter are the two pages to scan for extended-range alternatives.
Specifications
| Level | 0.01 |
|---|---|
| Temperature range | 0°C to 80°C |
| Stability | (±0.05pH±1 digit)/3h |
| Measuring Range | { |
| Resolution | { |
| Basic Tolerance | { |
| Electronic Unit Input Resistance | 1012 Ω |
| Electronic Unit Input Current | 1012 A |
| Power Source | Digital: DC (12V, 10A) |
| Dimension (W×D×H) | 290×210×95 mm |
| Net Weight | 1.5 kg |
Features
| Feature Area | Description |
| Backlit Display | A large backlit LCD shows pH, mV, and temperature readings together. |
| Temperature Compensation | Compensation can run automatically or be set by hand for the sample at hand. |
| Calibration | Multi-point calibration comes with automatic detection of the standard buffer solutions. |
| Auto-Hold | Auto-hold senses the measurement endpoint and locks the reading. |
| Keypad and Reset | A keypad with menu-specific function keys speeds navigation, and a reset returns all settings to factory defaults. |
Applications