Description
EziLab offers the Single Beam UV-Vis Spectrophotometer EZL-SBS104, a UV vis spectrometer covering 190 to 1100 nm at a 1.0 nm spectral bandwidth. Wavelength holds within 0.1 nm with repeatability of 0.1 nm or less, photometric readings stay within 0.3% T, and stray light measures 0.05% T at 220 and 360 nm. Deuterium and tungsten lamps feed a silicon photodiode over a thick optical base, drift holds at 0.001 A per 30 minutes, and a 320 by 240 dot LCD with USB and parallel ports completes the unit. That narrow 1.0 nm bandwidth resolves closely spaced spectral peaks that a 2 nm instrument can blur together. Automatic calibration, absorbance, transmittance, and concentration modes, plus bundled PC software for spectral scanning and kinetics, cover routine and method work. Four glass and two quartz cuvettes ship standard, micro cuvettes optional. Pharmaceutical QC labs confirm raw material identity and assay results, while food testing labs analyze chemical composition and additives. Environmental labs measure water parameters, life science groups quantify DNA and proteins, and universities teach with this UV spectrophotometer. Its 110 or 220 V input suits US benches. Deuterium lamps are the consumable to price: ask for rated hours and replacement cost, then compare the Spectrophotometer category and Single Beam UV-Visible Spectrophotometer range.
Specifications
| Wavelength Range | 190 to 1100 nm |
|---|---|
| Spectral Bandwidth | 1.0 nm |
| Wavelength Accuracy | ± 0.1 nm, ± 0.3 nm |
| Wavelength Repeatability | Less than or equal to 0.1 nm |
| Photometric Accuracy | ± 0.3 % T (0 to 100 % T) |
| Photometric Repeatability | Less than or equal to 0.1 % T (0 to 100 % T) |
| Photometric Range | - 0.3 to 3 (Absorbance), 0 to 200 % (Transmittance), 0 to 9999 (Concentration) |
| Stability | Less than or equal to 0.001 A / 30 min at 500 nm |
| Stray Light | 0.05 % T at 220 nm & 360 nm |
| Output | USB Port & Parallel Port |
| Display | 320 × 240 Dots LCD |
| Light Source | Tungsten Lamp & Deuterium Lamp |
| Detector | Silicon Photodiode |
| Power Requirements | AC 110 / 220 V, 50 / 60 Hz |
Features
| Feature Area | Description |
| Optical Base | Thick optical base carrying the optical parts. |
| Measurement Modes | Multiple measurement modes for absorbance, transmittance and concentration work. |
| Data Access | Quick access to stored data for review at the bench. |
| Dual Lamp Source | Tungsten and deuterium lamp pair covering visible and UV measurements. |
| PC Link | USB and parallel outputs, with PC software supplied. |
Standard Accessories
Spectrophotometer | 1 unit |
PC Software | 1 set |
1 cm Glass Cuvette | 4 pieces |
1 cm Quartz Cuvette | 2 pieces |
Power Cord | 1 piece |
User’s Manual | 1 piece |
Dust Cover | 1 piece |
Optional Accessories
1 | Cell Holders |
2 | Control Units |
3 | Lamps |
4 | Glass Cuvette with Lid |
5 | Quartz Cuvette with Lid |
6 | Micro Cuvette |
Applications