EziLab UV-Visible Spectrophotometer EZL-US132 performs absorbance, transmittance, and concentration measurements across pharmaceutical analysis, environmental monitoring, food quality testing, and chemical laboratory applications. Wavelength range spans 190 to 1100 nm with 2 nm spectral bandwidth and 0.5 nm accuracy for able UV spectrophotometer performance. Double beam optical system with silicon photodiode detector and dual deuterium and tungsten halogen lamps maintains stability within 0.001 A per hour, while stray light stays below 0.05% T at 220 and 360 nm. A color touchscreen display and four-cell holder streamline operation and parallel sample handling. Researchers and analysts can count on EziLab's Lab Equipment and spectrophotometers for a complete UV spectrophotometer and double beam built for precision.



Wavelength Range 190 nm – 1100 nm
Spectral Bandwidth 2 nm
Optical System Double beam
Wavelength Accuracy ±0.5 nm
Wavelength Repeatability ≤0.2 nm
Photometric Range -0.4 – 4A, 0 – 200%T, 0 – 9999C
Photometric Accuracy ±0.2%T (0-100%T)
±0.002Abs (0-0.5Abs)
±0.004Abs (0.5-1.0Abs)
Photometric Repeatability ≤0.15%T (0-100%T)
0.001Abs (0-0.5Abs)
0.002Abs (0.5-1.0Abs)
Stability ±0.001A/h @ 500 nm
Baseline Flatness ±0.001A
Stray Light ≤0.05%T @ 220 nm, 360 nm
Noise ±0.0005A
Cell Holder 4 cell holder
Display 800×840 color touch screen
Lamps Deuterium and Tungsten Halogen lamps
Detector Silicon photodiode
Data Output Port USB
Printer Port Parallel port
Power Supply AC220/50Hz, 110V/60Hz
Packing Dimension (W×D×H) 860×660×465 mm
Gross Weight 35 Kg
  • 800×840 color touch screen display

  • Stand-alone system which can recognize various measurements

  • Uses a rigid 8 mm aluminum base as optical mount to ensure stability and reliability

  • Automatic system calibration

  • Optical spare parts to ensure low stray light, stability and reliability

  • Data can be exported to USB directly

  • Intelligent alarm function

  • Displays several groups of data at same time for direct reading

  • System encryption function

  • Simple and elegant design for easy operation

  • Reduced handling steps

  • It can connect to the printer and output the measurement results directly

  • It is ideal for quality control, general research, pharmaceutical, biochemical and clinical laboratory applications. It is also extensively used for quality assurance tests in various educational institutes, universities and laboratories.