EziLab offers the Kinematic Viscometer EZL-KV100, a viscosity tester that holds a 20 L bath from ambient to 100°C within 0.01°C using a Pt100 sensor and a 1200 rpm stirrer. Four samples run at once, and the timer counts from 0.0 to 9999.9 seconds within 0.05% over 60 minutes. Six glass capillary viscometers with bores from 0.6 to 2.0 mm, capillary holders, thermometers, and stoppers come standard, and presettable viscosity coefficients turn flow time straight into a result. Analysts fill a capillary, let it settle in the bath, and time the sample between two marks, following ASTM D445. Stored coefficients and automatic printing remove manual math, so viscosity measurement stays consistent between shifts and operators. Tight bath control matters because kinematic viscosity shifts quickly with temperature, especially near 40°C and 100°C test points. Petroleum refineries and lubricant blenders grade incoming base oils, fuels, and finished lubricants, while used oil labs track machinery wear. Chemical producers check resins and solvents, and pharmaceutical, cosmetic, food, and beverage plants test syrups, creams, lotions, and liquid formulations. Contract labs value this viscometer for routine certification work, and university labs use it for teaching. A spec that lists both 40°C and 100°C means two bath runs, because the bath holds one setpoint at a time, so tally your monthly samples by test temperature before comparing the Petroleum Testing Equipment category and Kinematic Viscometer range.



Quantity:
$5992
Temperature Range Ambient to 100 ℃
Operating Temperature 15℃ to 35℃
Temperature Control Accuracy ± 0.01 °C
Relative Humidity ≤ 85 %
Timing Range 0.0 s to 9999.9 s
Timing Accuracy ≤ ± 0.05 % within 60 min
Stirring Motor 1200 rpm
Sample Quantity 4 piece at the same time
Bath Capacity 20 L
Temperature Sensor RTD, Pt100
Capillary Viscometer Tubes 6 pieces
Inner Capillary Viscometer Diameters 0.6mm, 0.8mm, 1.0mm, 1.2mm, 1.5mm, 2.0mm
Power Consumption ≤1800W
Dimensions 530 × 400 × 670mm
Weight 42kg
Feature AreaDescription
Result PrintingAutomatic printing of test results at the end of a run.
Viscosity CoefficientsPresettable viscosity coefficients entered before a run starts.
Temperature BathHeated bath with a stirring motor and RTD sensor to hold the test temperature.
Sample HandlingHolders for multiple capillary viscometers so several samples run together.
GlasswareGlass capillary viscometer tubes in a range of bore sizes come with the unit.
  • Capillary viscometer (total of 7) × 1 Set Silica gel plug (3#) × 1 Piece Capillary holder × 4 Piece Rubber tube (Φ10, 0.5m) Rod type mercury thermometer (0℃ to 50℃, 50℃ to 100℃, scale divisions are both 0.1℃) ×1 Piece Standard Second Thermometer ×1 Piece Cover A × 4 piece Stopping ring for thermometer (1#) × 2 Piece 8.15A fuse (Ф5 × 20) × 2 Piece

  • Supports flow time checks on base oils, fuels and finished lubricants in refinery and lubricant blending labs.
  • Handles viscosity change tracking on service samples in used-oil analysis labs.
  • Suited to comparing resin and solvent batches in chemical plant labs.
  • Applied to flow consistency testing of liquid formulations in pharmaceutical, cosmetic, food and beverage labs.
  • Employed for routine contract lab work with automatic result printing.
  • Fits capillary bore selection to suit the thickness of the sample.