EziLab offers the Gasoline Oxidation Stability Tester EZL-OST51, a fuel tester that measures the induction period of gasoline, the time before a sample begins rapid oxidation under heat and pressure. That figure predicts how long fuel stays stable in storage before gum and deposits form. Choose it over the lubricating oil oxidation tester when the sample is motor gasoline or a blending component, because the two methods use different conditions and answer different questions: storage life for fuel versus service breakdown for oil. A metal water bath holds 100°C within ±0.5°C on 1600 W of heating, with efficient insulation keeping the temperature steady through long runs. Pressure reads from 0 to 1600 kPa with ±2‰ accuracy, and automatic detection of the oxidation turning point marks the end of the induction period without operator guesswork. A dual-hole layout tests two samples side by side for duplicate runs or batch comparisons. A coefficient-adjustable mercury-in-glass thermometer provides a reference check. Petroleum refineries verify gasoline storage stability before release, fuel testing laboratories certify blends and additives, and automotive research teams study how fuel aging affects engine deposits. Each gasoline stability method fixes a bath temperature and pressure window, so match your lab’s method sheet to 100°C and 0 to 1600 kPa before comparing the Petroleum Testing Equipment category and Oxidation Stability Tester range.



Quantity:
$11429
Temperature Control 100 ℃
Temperature Control Accuracy ± 0.5℃
Heating Power 1600W
Measuring Range of Pressure Transmitter 0 to 1600 kPa
Accuracy ± 2‰
Thermometer Mercury in a glass thermometer can correct the coefficient as need.
Ambient Temperature 15℃ to 35℃
Relative Humidity ≤ 85%
Power Consumption ≤ 1400 W
Dimension (L×W×H) 470×380×320 mm (without test barrel)
470×380×600 mm (with test barrel)
Weight 25 kg
Feature AreaDescription
Turning PointDetects sample oxidation turning points.
Dual-Hole BathDual-hole layout for testing two samples together.
InsulationHeat preservation and insulation system.
Water BathBuilt-in metal water bath.
PressurePressure transmitter for reading pressure during the test.
  • Supports verifying gasoline storage stability before release in petroleum refineries.
  • Handles testing of gasoline blends and additives in fuel testing laboratories.
  • Suited to studying how fuel aging affects engine deposits in automotive research.
  • Applied to measuring the induction period of motor gasoline and blending components.
  • Employed for comparing two gasoline samples side by side in the dual-hole bath.
  • Fits reference temperature checks with the mercury thermometer.